Protocols: Acid Cleaning Coverslips
-
Make-up 300 ml of 2 parts Nitric Acid - 1 part HCl in a glass
beaker in the hood. Solution will turn orange-red.
-
Place #1.5 coverslips into solution a few at a time. Let sit
for about 2 hrs. with occasional swirl.
-
Decant acid into waste bottle carefully.
-
Wash extensively in running water until pH of water is back
up to 5.5 -6.0.
-
Store in 70% EtOH.
-
Flame coverslip in hood prior to use.
Return